Anesthesia for Correction of Cardiac Arrhythmias
Samuel A. Irefin
Key Points
|
|
1. |
Cardiac arrhythmias are caused by disorders of impulse formation or disorders of impulse conduction, or both. Cardiac arrhythmia may be life-threatening because of a reduction in cardiac output, reduction in myocardial blood flow, or precipitation of a more serious arrhythmia. |
|
|
2. |
Radiofrequency ablation is the therapy of choice for many types of cardiac arrhythmias. |
|
|
3. |
Electrophysiologic studies are used to map out normal and abnormal intracardiac structures. In this process, the mechanism of arrhythmia is delineated, and ablation can be performed at the same time. |
|
|
4. |
Pacing technologies have been developed to treat heart failure resulting in increases in pulse pressure, left ventricular stroke volume, cardiac index, and wedge pressure. |
|
|
5. |
Implantable pacemakers are placed for treatment of symptomatic bradycardia with the ability to respond to changing hemodynamic demands. |
|
|
6. |
The development of implantable cardioverter-defibrillators (ICDs) to terminate ventricular tachyarrhythmias by delivering high-voltage shocks to the ventricle has revolutionized therapy for cardiac arrhythmias. |
|
|
7. |
The main purpose of ICD placement is to prevent sudden cardiac death resulting from hemodynamically unstable ventricular arrhythmias. |
|
|
8. |
An ICD also can be placed for cardiac resynchronization. Cardiac resynchronization therapy has been shown to improve heart failure symptoms, quality of life, exercise capacity, and electrocardiographic variables. |
|
|
9. |
Anesthetic management of patients for correction of cardiac arrhythmias depends on associated comorbid illness and the procedure that is planned. |
Cardiac arrhythmias are caused by disorders of impulse formation or disorders of impulse conduction, or both. Disorders of impulse formation include enhancement or depression of automaticity, parasystolic activity, and triggered activity. Disorders of conduction include decremental conduction, re-entry, entry block, exit block, concealed conduction, and supernormal conduction.[1]
At the present time, radiofrequency catheter ablation has replaced antiarrhythmic drug therapy as the treatment of choice for many types of cardiac arrhythmias. Before the 1980s, cardiac electrophysiology was primarily used to confirm mechanisms of arrhythmias, with management mainly by pharmacologic means. As a result of shortcomings in antiarrhythmic drug therapy (including the results of randomized trials), radiofrequency ablation and implantable cardioverter-defibrillators (ICDs) were developed. [2] [3]
Historical Perspectives
The treatment of cardiac arrhythmias with device-based therapy may have begun in 1899, when Prevost and Batteli[4] noted almost as an afterthought that direct electric shock could terminate ventricular fibrillation in dogs. Hooker and colleagues[5] showed 3 decades later that the passage of electric current across the heart can initiate and terminate ventricular fibrillation. In 1947, Beck saved the first human life by the successful use of cardiac defibrillation in a 14-year-old boy, who developed ventricular fibrillation during a thoracic procedure and went on to achieve full recovery.[6] These early achievements provided the foundation for the landmark work of Mirowski and Mower,[6a] which ultimately led to the development of ICDs in humans in 1980.
Scope of Cardiac Arrhythmias
Cardiac arrhythmias are common (see Chapters 42 and 43 [Chapter 42] [Chapter 43] ). Some cardiac arrhythmias are life-threatening, and others are merely a nuisance. Cardiac arrhythmias are caused by abnormalities in impulse formation or conduction that lead to slow or fast, regular or irregular heart rhythms. At the present time, it is not difficult to treat slow rhythms because available pacemakers are able to adapt slow function to the needs of the body.[7] The situation is different, however, for patients with rapid rhythms. Rapid rhythms may originate anywhere in the heart and result from various mechanisms. These mechanisms may be focal, meaning that the abnormal impulse formation is confined to a small area, or they may be the result of an impulse running in a circuit composed of several interconnected cardiac cells. Such a circuit may be small or large, as in atrial flutter and in arrhythmias in which the normal atrioventricular conduction system and an extra connection between the atrium and the ventricle are incorporated into the circuit of the arrhythmia.[8]
Pharmacologic interventions originally were used in attempts to terminate and prevent rapid rhythms. It has become clear in the past several decades, however, that antiarrhythmic drugs may have serious side effects and sometimes may even facilitate the occurrence of life-threatening arrhythmias and sudden death.[9] As a result of these effects, techniques were developed for localizing the site of origin or pathway of an arrhythmia and then isolating or destroying the tissue that is responsible. By employing an intracardiac catheter, it is now possible to determine the site of origin or pathway of an arrhythmia and cure the rhythm disturbance by applying through the catheter radiofrequency, laser, ultrasound, or microwave energy or freezing temperatures to the tissue causing the arrhythmia.
Heart failure is a major problem in elderly patients (see Chapter 71 ). Although pharmacologic treatment of heart failure has improved, outcome generally remains poor. New pacing technologies may now be used to treat selected patients with heart failure. For many years, permanent pacing has been used to treat symptomatic bradycardia, and pacing may alleviate heart failure when associated with heart block. Several studies have examined the use of conventional dual-chamber atrioventricular–right ventricular pacing for treatment of heart failure in the absence of symptomatic bradycardia or heart block. [3] [10] Biventricular pacing aims to restore synchronous cardiac contraction. Studies have shown that when ventricular desynchrony is reduced, the heart is able to contract more efficiently and increase left ventricular ejection fraction and cardiac output, while working less and consuming less oxygen.[11] In addition, re-establishment of left ventricular synchrony can increase left ventricular filling times, decrease pulmonary capillary wedge pressure, and reduce mitral regurgitation.
Normal Cardiac Rhythm
In the normal heart, the dominant impulse arises in the sinus node with a rate of 60 to 100 beats/min. During sleep, the rate may decrease to 30 to 50 beats/min.[12] Episodes of sinus pauses up to 3 seconds, sinoatrial block, junctional rhythms, and first-degree and second-degree atrioventricular nodal block that occur often enough (especially in trained athletes) are considered to be normal variants.[7]
The impulses generated from the sinoatrial node propagate along three intra-atrial conduction pathways: the anterior, middle, and posterior internodal tracts. These tracts are not discrete pathways, but groups of cells that conduct slightly faster than the atrial myocardium.[13] The internodal tracts give rise to interatrial fibers. The electric impulse, whether propagated in the atrial myocardium or along the internodal tracts, converges on the atrioventricular junction. The atrioventricular node located in the atrioventricular junction ultimately receives the impulses generated from the sinoatrial node. The impulses are delayed in the atrioventricular node before they are finally distributed to the ventricular myocardium via the His-Purkinje system.
Normally, the heart rate increases with exercise to at least 85% of the age-predicted maximum of 220 minus age in years; failure to do so is termed chronotropic incompetence. Sinus arrhythmia is defined as sinus rhythm with P-to-P variations of more than 10%. Sinus arrhythmia is due to cyclic variations in vagal tone commonly related to respiration (the rate is faster with inspiration and slower with expiration).[14] Sinus arrhythmia disappears with exercise, breath-holding, and atropine, and is more likely to be seen in individuals who do not have heart disease.[15]
Cardiac Arrhythmias
Cardiac arrhythmia is caused by a disorder of impulse generation, impulse conduction, or a combination of both. Cardiac arrhythmia may be life-threatening because of a reduction in cardiac output, reduction in myocardial blood flow, or precipitation of a more serious arrhythmia.[16] Arrhythmias may be described based on (1) rate (bradycardia or tachycardia); (2) rhythm (regular or irregular); (3) origin of impulse (supraventricular, ventricular, or artificial pacemaker); (4) impulse conduction (atrioventricular, ventriculoatrial, or block); (5) ventricular rate; or (6) special phenomena (e.g., pre-excitation).
Re-entry is a common electrophysiologic mechanism that predisposes to most ventricular arrhythmias and to most supraventricular tachyarrhythmias. The most common mechanism of re-entry is based on the model originally proposed by Erlanger and Schmitt and later modified by Wit.[1] This model postulates the presence of a ring or loop of cardiac tissue that is functionally separate from neighboring tissue and the presence of transient or permanent unidirectional block in a portion of the loop. Unidirectional block may be anatomic in origin (e.g., bundle branches, fibrosis, dual pathways, atrioventricular node plus accessory pathway) or functional (e.g., ischemia, drug effect).
Atrial flutter is a macro-re-entrant arrhythmia identified by flutter waves, often best seen in the inferior leads at 250 to 350 beats/min. Patients often present with a 2 : 1 atrioventricular conduction with a ventricular rate of 150 beats/min, although the atrioventricular conduction ratio can change abruptly.
Atrial fibrillation is a narrow-complex tachyarrhythmia and is the most common in the general population. The prevalence of atrial fibrillation in the general population increases exponentially with age, from 0.9% in individuals age 40 to 5.9% in individuals older than age 65. The most important risk factors for development of atrial fibrillation in the general population are structural heart disease, valvular heart disease, and left ventricular hypertrophy.[17]
Ventricular tachyarrhythmia is defined as three or more consecutive ectopic beats at a rate greater than 100 beats/min.[18] Ventricular tachyarrhythmia is traditionally classified as nonsustained or sustained. Sustained ventricular tachyarrhythmia is defined as ventricular tachyarrhythmia lasting more than 30 seconds. Nonsustained ventricular tachyarrhythmia is defined as ventricular tachyarrhythmia that terminates spontaneously within 30 seconds. Sustained ventricular tachyarrhythmia also is traditionally classified as monomorphic (one site of origin) or polymorphic (two or more sites of origin).[18] Monomorphic ventricular tachyarrhythmia usually results from re-entry, and the site of re-entry depends in part on the type of heart disease. In patients with coronary artery disease, the re-entry circuit is usually located in ventricular myocardium, whereas in dilated cardiomyopathy with left bundle branch block, bundle branch re-entry is common.[19] Monomorphic ventricular tachyarrhythmia may occur in individuals with an otherwise normal heart, whereas polymorphic ventricular tachyarrhythmia may occur in acquired states that produce a marked prolongation of the Q–T interval. Nonsustained ventricular tachyarrhythmia is frequently asymptomatic, but may produce palpitations, weakness, and presyncope.[19]
Torsades de pointes is a French term translated as “twisting of the points.” It is a syndrome composed of polymorphic ventricular tachyarrhythmia. It may be due to various medications or electrolyte imbalances. Torsades de pointes is usually paroxysmal, but is frequently symptomatic and often produces loss of consciousness. It occasionally degenerates to ventricular fibrillation.
Ventricular fibrillation accounts for 80% to 85% of sudden cardiac deaths.[19] Ventricular fibrillation is usually preceded by ventricular tachyarrhythmia, but also may occur as a primary arrhythmia. More recent studies suggest that ventricular fibrillation results from multiple wavelengths that disperse randomly, using the leading circle form of re-entry.[19] The most common cause of ventricular fibrillation is acute myocardial infarction. It also is observed in patients with chronic ischemic heart disease, hypoxia resulting from any cause, acidosis, hypokalemia, and massive hemorrhage.
Indications for Correction of Cardiac Arrhythmias
Intracardiac electrophysiologic studies can give valuable information about normal and abnormal electrophysiology of intracardiac structures (see Chapters 42, 43, and 97 [Chapter 42] [Chapter 43] [Chapter 97] ). These studies are used to confirm the mechanism of an arrhythmia, to delineate its anatomic substrate, and to ablate it. The electric stability of the ventricles also can be assessed, as can the effects of an antiarrhythmic regimen.
In addition, pacing technologies have been developed to treat heart failure with promising results, leading to improvement in morbidity and mortality in these patients. Hemodynamic responses to biventricular pacing include an increase in the rate of elevation of left ventricular pressure and increases in pulse pressure, left ventricular stroke work, cardiac index, and wedge pressure.[20] Cardiac resynchronization therapy improves ventricular function without increasing myocardial energy consumption, in contrast to the effect of inotropic agents, such as dobutamine.[11] In addition, cardiac resynchronization therapy may reverse left ventricular remodeling over time.[21]
Permanent Pacing
Indications for pacemaker therapy have increased in recent years and now include the treatment of bradyarrhythmias and heart failure according to the American College of Cardiology and American Heart Association guidelines.[22] These guidelines discuss indications for pacing in patients with sinus node dysfunction, acquired atrioventricular block, chronic bifascicular and trifascicular block, hypersensitive carotid sinus, and neurally mediated syndromes. The guidelines serve to direct the treating physician in selecting which patients would benefit from device therapy.
A Swedish team led by Sennings and Elmqvist implanted the first pacemaker in 1958.[23] A thoracotomy was required, and pacing was done through electrodes sutured to the epicardium. In these early systems, significant problems with changes in pacing threshold, lead infection, and lead breakage were common. Transvenous lead implantation subsequently developed by Furman would resolve many of these issues.[24] In 1958, Furman successfully paced an elderly patient with a catheter electrode inserted transvenously. Other investigators took on the challenge of solving various technical problems, such as device miniaturization, longer life batteries, and stable, reliable lead material.[25]
As the indication for implantation expanded from atrioventricular conduction disturbances to management of sinus node dysfunction, the need for implantable pacemakers grew in proportion.[25] Technology evolved rapidly with the development of lithium-iodide batteries that had greater longevity. Electronic advances then led to major miniaturization using integrated circuits as opposed to discrete components. Lead materials used in today's pacemaker rely on silicone and polyurethane, which are more biocompatible and reliable than earlier materials. With all these technical refinements, present-day pacemakers are small and can pace reliably for 8 to 10 years before generator replacement is needed.
The primary functional challenge for contemporary pacemakers is to maintain the heart rate based on circulatory needs, pacing in a manner that mimics the natural physiology of excitation and conduction. In a healthy heart, the sinus node is modulated by the autonomic nervous system, and its rate is determined by a multiplicity of factors, such as physical activity, emotion, and blood pressure. Not only the rate, but also the activation sequence and atrioventricular conduction time vary with demand; these requirements also must be considered. Rate is controlled by pacemaker discharge, and the excitation and conduction sequence depend on the placement of pacing electrodes.
Approximately 120,000 pacemakers are implanted each year in the United States. Indication for implantation for most of these cases is sick sinus syndrome. Other indications include atrioventricular block, carotid sinus hypersensitivity, malignant vasodepressor syndrome, and hypertrophic cardiomyopathy.[26]
The primary purpose of implantable pacemakers is to treat symptomatic bradycardia. With the extraordinary developments that have occurred in pacemaker therapy for the traditional indication, bradycardia, new uses are now beginning to be explored. Pacemakers have progressed from large, fixed-rate, single-chamber devices to multiprogrammable, multichamber devices with the ability to respond to changing hemodynamic demands. As technology advances, other possible uses are likely to be conceived.
Resynchronization Therapy
Cardiac resynchronization therapy has been shown to improve heart failure symptoms, quality of life, exercise capacity, hospitalization, and echocardiographic variables.[27] Based on the available data, cardiac resynchronization therapy is indicated in patients with drug-refractory, symptomatic New York Heart Association functional class III and IV heart failure of either ischemic or nonischemic origin.[28] In addition, these patients are protected from associated risk of sudden cardiac death when combined with an ICD system.[29]
The development of an automatic internal defibrillator or ICD began in the 1960s. External cardiac defibrillation was increasingly being used in coronary care units for the treatment of ventricular fibrillation and sudden cardiac death. Although the idea of automatic external defibrillation had been discussed initially by Zycoto, Mirowski and colleagues[30] were the first to champion and begin practical development of an automatic internal device. In 1969, Mirowski and Mower developed the prototype of today's automatic internal defibrillator.[31]
The primary goal of all defibrillators is to terminate ventricular tachyarrhythmias by delivering high-voltage shocks to the ventricle. As with implantable pacemakers, defibrillating devices need to be small and reliable, and to have adequate longevity. ICDs have evolved not only to perform this function, but also to take on additional tasks, such as antitachycardia pacing of the ventricle, dual-chamber pacing, and termination of atrial tachyarrhythmias.
A key difference between pacing and defibrillation of the heart is that for pacing; only a very small mass of myocardium needs to be stimulated, whereas for defibrillation, most, if not all, of the myocardium must be stimulated. Because the myocardium is easily excitable throughout diastole, a small wave of depolarization during pacing can readily propagate throughout the whole heart. In contrast, during ventricular fibrillation, there are usually multiple re-entrant wave fronts that are continuously changing in location and size that must be quelled. To defibrillate successfully, most of these wavefronts have to be interrupted simultaneously; to achieve this, one needs to capture most of the tissue that is in a state of relative refractoriness.[32] One unique property of defibrillation success is that it is probabilistic.[33] The same energy that can defibrillate the heart on one occasion may be unsuccessful at another time.
The main purpose of ICD placement is to prevent death from hemodynamically unstable ventricular tachyarrhythmias. Although advances in technology have made these devices much more flexible in terms of arrhythmia detection and electric therapy potions, their main purpose is to reduce sudden cardiac death, which claims approximately 300,000 lives in the United States annually. Secondary prevention of sudden cardiac death in patients who have survived cardiac arrest is another major indication for ICD placement. In such patients and especially in patients of this group for whom no reversible or curable cause can be found, ICD implantation has been repeatedly documented to provide a major mortality benefit.[34] Interest in managing atrial tachyarrhythmias also has grown significantly in recent years. It is now recognized that about 30% of patients with ventricular tachyarrhythmia also have atrial tachyarrhythmias.[35] Such atrially initiated tachyarrhythmias can worsen patient symptoms, can result in inappropriate ventricular shocks, and may be responsible for initiating ventricular tachyarrhythmias that can exacerbate other pathologies, such as heart failure. New strategies for treatment and prevention of atrial tachyarrhythmias are incorporated into devices that are capable of defibrillation and antitachycardia pacing in the atrium and the ventricle, in addition to combined dual-chamber pacing.[36]
The relative ease of ICD implantation and longevity of current defibrillators have made them a valuable tool in primary prevention. Patients no longer must survive a cardiac arrest to justify the risk of ICD implantation.
Preoperative Evaluation
Most patients who require pacemaker or ICD placement have significant cardiovascular disease. In addition, correction of cardiac arrhythmia may require radiofrequency catheter ablation. Radiofrequency catheter ablation has proved highly effective in the treatment of atrioventricular nodal re-entrant and accessory pathway tachycardias. Indications for pacemaker and ICD placement continue to evolve as the utility of these devices continues to increase. Although most pacemaker placement is done with local anesthetic infiltration, ICD placement may require monitored anesthesia care or in some cases general anesthesia. The modern ICD unit is capable of delivering the full spectrum of therapy for ventricular tachyarrhythmias and for bradycardia therapy with dual-chamber pacing/sensing, rate modulation, and mode-switching features.
As mentioned earlier, there are two common indications for ICD placement. One is continued ventricular tachyarrhythmias despite adequate drug therapy. Another indication is history of sudden cardiac arrest that is not associated with myocardial infarction.
Preoperative evaluation processes necessary for placement of an ICD should be complete by the time the decision is made to place the device (see Chapters 34 and 35 [Chapter 34] [Chapter 35] ). These patients need a thorough preoperative evaluation. This evaluation includes electrophysiologic testing to determine the inducibility of ventricular tachycardia and electrophysiologically guided drug therapy. Preoperative pulmonary function tests may be necessary in patients on amiodarone to evaluate possible toxicity of this drug, which can result in chronic obstructive pulmonary disease or interstitial lung disease. In some instances, the underlying pathophysiology of malignant ventricular arrhythmias is related to ischemic or idiopathic cardiomyopathy.[37] These patients often present with poor left ventricular function and higher incidence of congestive heart failure. Patients with a history of congestive heart failure should be in optimal condition before surgery.
Generally, all patients who present for correction of cardiac arrhythmia require preoperative evaluations including electrocardiogram, chest radiograph, hemoglobin, and electrolytes. Patients should be NPO (have nothing by mouth) at least 8 hours before the procedure. In addition, patients who require device and lead extractions because of malfunction or infection may require blood product transfusions during the procedure. Consequently, type and crossmatch of blood products is frequently necessary for these procedures.
Anesthetic Considerations
Pacemakers
Permanent implantable pacemakers have been the standard modality of treatment for patients with all types of bradyarrhythmias. A significant number of these patients present with sick sinus syndrome and are older. Consequently, devices are placed under general anesthesia in these patients. As a result of more recent advances in pacemaker technology, these devices now can be placed as a therapeutic modality to alter hemodynamic states. Surgeons used to be primarily responsible for device insertion. Now the task falls under the services of cardiologists. Device placement is commonly performed in the cardiac catheterization suite under local anesthesia on an outpatient basis. Complicated high-risk patients now present for pacemaker insertion, however, in addition to more recent indications by the American College of Cardiology/American Heart Association for these devices. In light of these increased indications, the expertise of anesthesiologists is needed for monitoring and perioperative care of these patients.
Monitored Anesthesia Care
Currently, most pacemaker insertions are performed by cardiologists. Most of these cases are performed under local anesthesia with sedation. Depending on the level of training, administration of sedatives and analgesics can be provided by nurses.
In instances that require deeper sedation for a patient's comfort or for critically ill patients with hemodynamic instability, monitored anesthesia care by an anesthesiologist may be required). Adequate monitoring and resuscitation equipment is required in such situations. The goal of monitored anesthesia care is to provide analgesia, sedation, and anxiolysis, while ensuring rapid recovery with minimal or no side effects. Any sedative-hypnotic medication may be used during monitored anesthesia care with a wide variety of delivery systems.[38] Subanesthetic concentrations of inhaled agents also have been used to supplement local anesthetics. Newer drugs, such as centrally mediated α2-agonists, have been shown to produce anxiolysis, sedation, and reduced requirements for supplemental analgesic medications during monitored anesthesia care.
General Anesthesia
Patients requiring pacemaker placement rarely require general anesthesia for placement. If general anesthesia is required, it should be directed toward underlying cardiac pathophysiology, indications, complications, and hemodynamic goals. Immediate access to life-support equipment, such as a cardiac defibrillator and a transcutaneous pacemaker, is necessary if the device is being placed under general anesthesia.
Implantable Cardioverter-Defibrillator
Since the 1980s, indications for use and implantation of ICDs have steadily increased. Over the past 2 decades, ICDs have undergone a significant evolution. In the 1970s and 1980s, ICD placement usually required thoracotomy for placement of epicardial patches.
Preoperative Evaluation
As mentioned earlier, common indications for ICD implantation include continued ventricular tachyarrhythmias unresponsive to adequate pharmacotherapy and history of sudden cardiac arrest unassociated with myocardial infarction. Newer indications include patients with various forms of the congenital long QT syndrome.[39] Patients with long QT syndrome who have already survived an episode of cardiac arrest or documented polymorphic ventricular tachyarrhythmia, especially if on pharmacotherapy at the time, are increasingly being evaluated as ICD candidates. In addition, patients with hypertrophic cardiomyopathies and without a history of sudden death are usually evaluated for ICD placement.[40] In these patients, sustained ventricular arrhythmias, nonexertional syncope, or a strong family history of sudden death with early age of presentation strongly indicates ICD implantation.
In all instances, the evaluation that is necessary for ICD implantation is completed by the time the decision is made to place the device (see Chapters 34 and 35 [Chapter 34] [Chapter 35] ). Electrophysiologic studies may have been done to determine the forms of arrhythmias present. When the pathophysiology of ventricular arrhythmias is related to idiopathic or ischemic cardiomyopathy,[41] these patients may present with poor left ventricular function and a high incidence of congestive heart failure. Consequently, they should be optimized as much as possible preoperatively.
Anesthetic Considerations
In the 1980s, ICD implantation was done with epicardial leads via thoracotomy under general anesthesia with one-lung ventilation. The technologic development of implantable ICDs with transvenous lead systems has simplified their implantation. Consequently, it was reasoned that ICDs can be placed under deep sedation with little or no intervention by the anesthesiologist analogous to what is needed for pacemaker placement.[42] Placement of an ICD under general anesthesia may be safer and more comfortable for the patient, however. Patients who present for ICD placement are often critically ill with cardiopulmonary comorbidity. It is not unusual for these patients to present with ejection fractions less than 30% and to require vasopressors to support hemodynamics during the procedure. In addition, some form of general anesthesia is necessary for intraoperative testing of defibrillating thresholds.
Monitored Anesthesia Care
Small, new-generation devices and transvenous lead systems lend themselves to the use of local anesthesia and intravenous sedation for ICD implantation. Midazolam and fentanyl are usually the drugs of choice when an ICD is placed under monitored anesthesia care (see Chapter 78 ). Monitoring includes pulse oximetry, five-lead electrocardiogram, and noninvasive blood pressure. Depth of anesthesia is monitored clinically. One of the major aspects of ICD placement is testing the device. Testing the device may require deep sedation or general anesthesia because the shocks that are associated with this procedure can be very painful. The presence of an anesthesiology team may be necessary for ICD placement under monitored anesthesia care.
General Anesthesia
Most patients who present for ICD placement typically have comorbidities such as ventricular tachycardia, congestive heart failure with ejection fraction less than 30%, coronary artery disease, pulmonary hypertension, chronic renal insufficiency, or valvular heart disease. These patients may be unable to lie flat for the prolonged period necessary for placement of the ICD. In addition, they may require close hemodynamic monitoring during the testing of the device. General anesthesia should be considered in these patients. When general anesthesia is chosen, in addition to standard monitoring, an arterial line may be added. External cardioverter-defibrillator pads are required for all ICD placements. These are employed in cases where an implanted defibrillator fails. General anesthesia also may be requested for anxious and extremely nervous patients. Because pacemakers and ICDs are placed percutaneously, anesthesiologists must be vigilant to possible complications, such as myocardial infarction, stroke, possible cardiac injury (perforation/tamponade), and pneumothorax from subclavian vascular access.
Extraction of Devices
As a result of continued growth and expanding indications for pacemakers and ICD placement, leads may require extraction because of mechanical dysfunction, the need to upgrade to more complex devices, or local or systemic infection. Lead extractions are probably one of the most challenging procedures that a cardiac electrophysiologist has to face today.
Indications for lead extractions can be divided into two categories—patient-related and lead-related. Patient-related indications include infection, ineffective therapy (high defibrillation threshold), perforation, migration, embolization, induction of arrhythmias, venous thrombosis, unrelenting pain, device interactions, and device upgrades.[43] Lead-related indications include lead recalls, lead failure, and lead interactions.[44]
Lead extraction is performed via powered sheaths through which energy is delivered to the tip in the form of excimer laser light or electrocautery. These systems burn through scar tissue adherent to the wall of the lead throughout its course. The potential for life-threatening complications, such as lead fracture, venous or myocardial rupture, and tamponade, makes general anesthesia with invasive monitors a prudent choice for lead extractions.
Postoperative Care
Postoperative care of patients with pacemaker or ICD implantation depends on various factors surrounding the implantation of the device (see Chapter 85 ). As mentioned earlier, most of these patients are quite ill with significant comorbidities. It is not unusual for patients to have congestive hart failure with an ejection fraction less than 30% as a result of poor left ventricular function. Consequently, it is imperative to have these patients monitored in the postanesthesia care unit, especially if the device is placed or extracted under general anesthesia. The spectrum of recovery sites after these procedures may vary from postprocedure units to a coronary intensive care unit. Most of these procedures are done on an outpatient basis; anesthesia is tailored to ensure rapid recovery after implantation.
Correction of Cardiac Arrhythmias with Ablation Therapy
Catheter ablation is a safe and curative option for most cardiac arrhythmias, with 85% to 98% cure rates among the arrhythmias treated most frequently.[45] The rate of major complications is less than 3%.[45] Cardiac ablation therapy involves the delivery of energy through a catheter that is usually placed in the endocardial position in the heart, destroying myocardial tissue that is responsible for the tachyarrhythmia. Multiple electrodes are inserted to locate the arrhythmia and ablate it. Usually the diagnostic portions of the ablation study are done during the same procedure.[46] The efficacy of catheter ablation depends on the accurate identification of the site of origin of the arrhythmia. When the site is identified, the electrode catheter is positioned in direct contact with the site of the arrhythmia, and radiofrequency energy is delivered through the catheter to destroy it.
The current that is generated by radiofrequency is alternating current, and is delivered at cycle lengths of 300 to 750 kHz when used for catheter ablation.[47] It causes resistive heating of the tissue in contact with the electrode. The degree of tissue heating is inversely proportional to the radius to the fourth power.[48] Consequently, the lesions created by radiofrequency energy are small. Although electric injury may be a contributing factor, the primary mechanism of tissue destruction by radiofrequency current is thermal injury. Acute lesions created by a radiofrequency current consist of a central zone of coagulation necrosis surrounded by a zone of hemorrhage and inflammation.[49]
Cardiac arrhythmias that can be treated with radiofrequency ablation include paroxysmal supraventricular tachycardia, Wolff-Parkinson-White syndrome, atrial flutter, atrial fibrillation, and idiopathic ventricular tachycardia. Most cardiac arrhythmias that are treated with radiofrequency ablation are not life-threatening, but have a significant impact on a patient's quality of life.[50] Advantages of radiofrequency ablation of cardiac arrhythmias include relief of symptoms, improvement in functional capacity and quality of life, and elimination of the need for lifelong antiarrhythmic drug therapy. The principal disadvantage is the risk of complications, which varies depending on the type of ablation procedure and skill of the operator.
Anesthetic Considerations
Catheter ablation was introduced into clinical practice in 1982. Initially, ablation was performed with direct electric shocks.[51] As a result of several advantages over direct current, radiofrequency ablation has replaced direct current ablation. These advantages include the absence of skeletal and cardiac muscle stimulation, minimal discomfort during delivery of energy, the possibility of performing the procedure in conscious patients, and the discrete nature of resulting lesions.[49]
Most cardiac ablation therapy for correction of arrhythmias can be performed under moderate sedation or monitored anesthesia care. In some of these cases, deep sedation may be required as the case progresses. In a few cases, general anesthesia may be required if the patient is anxious or cannot tolerate lying in the supine position for an extended period. General anesthesia may be implemented for these patients with standard American Society of Anesthesiologists monitors with adequate vascular access. Catheter ablation is the first-choice treatment for most cardiac arrhythmias. It is a safe treatment and is usually effective as a single procedure. Because it is curative in many patients, it is offered to all patients who would otherwise be committed to long-term drug therapy.
Future Trends
Correction of cardiac tachyarrhythmias has improved dramatically in the past 2 decades. Emphasis has shifted from pharmacologic therapy to nonpharmacologic therapy of tachyarrhythmias; this has led to a significant increase in the numbers of radiofrequency catheter ablations and defibrillator implantations. These developments were triggered by technologic advances that showed superiority of these procedures over the use of antiarrhythmic drugs.[52] As a result, treatment of supraventricular tachycardias and tachycardias involving accessory atrioventricular pathways will probably remain the domain of catheter ablation. The cure rates of patients treated with catheter ablation is very high. In addition, treatment of life-threatening ventricular tachyarrhythmia will remain in the domain of ICDs for the foreseeable future. The role of ICD therapy has been clearly defined with respect to prolongation of life and has been expanded to include primary prophylaxis of sudden death in high-risk populations.[53]
As a result of these developments, the presence of an anesthesiology team will continue to grow in cardiology suites. Patients who are being cared for in these areas are sicker with significant comorbidities. The role of conscious sedation will continue to diminish in the performance of these procedures. These patients will require full monitoring and care under the direction of an anesthesiologist.